普瑞森
  • Product: High efficiency and low-pass dialyzer

High efficiency and low-pass dialyzer
1. High-performance modified polysulfone membrane
2. Highbiocompatibility
3. High efficient small molecules clearance 
4. Adsorption of toxins and pyrogens
 
Type 
LO PS 10
LO PS 12
LO PS 15
LO PS 18
LO PS 20
Ultrafiltration Coefficient
6.8
7.9
9.8
11.3
12.3
Test Conditions of Urea Transport Area Parameters: QB = 200, QD = 500
562
570
809
911
1015
Effective Area (m2)
1
1.2
1.5
1.8
2
Preshoot Amount (ml) of Blood Side
58
68
90
104
113
Clearance Rate
QB 200ml/min
Urea
176
183
189
191
193
Creatinine
157
166
173
178
180
Phosphate
126
139
146
155
162
Vitamin B12
68
77
83
100
109
QB 300ml/min
Urea
217
233
246
251
256
Creatinine
181
200
213
219
225
Phosphate
147
162
172
187
198
Vitamin B12
73
82
91
112
125
QB 400ml/min
Urea
242
261
285
291
298
Creatinine
198
220
239
251
256
Phosphate
159
175
190
205
220
Vitamin B12
75
86
95
121
136
Wall Thickness/Inner Diameter (mm)
40/200
Membrane Material
Low-pass α polysulfone membrane (PVP/PS)
Casing
Polycarbonate
Sterilization
γ-ray
In vitro experiments and physical data comply with EN1283 (test conditions of clearance rate: QD = 500ml/min; QF = 0ml/min; ultrafiltration coefficient: human blood; hematocrit: 32%; total protein: 6%; T = 37 °C